- A prepared test for monitoring the cleaning efficiency of Flexible Endoscope-washer
- Two separate test soils for demonstrating the removal of blood and polysaccharides
- Dismountable device for visual check
- The device simulates a Flexible Endoscope channel
ENDOSCOPE WASHERS NEED TO BE REGULARLY MONITORED
The FlexiCheck-system is a unique product for monitoring the cleaning efficiency of the Flexible Endoscope washer. Flexible Endoscopes are often contaminated with mucus however blood is also a possible contaminant. FlexiCheck will monitor the efficiency of the washer and the detergent.
TOSI-FlexiCheck combines 2 different test soils on one test strip: One is the standardised blood test soil and the second is a polysaccharide mixture.
For reprocessing Flexible Endoscopes washers with channel irrigation systems are recommended. FlexiCheck is connected to this system and will simulate a single channel at one time.
EVALUATION

CLEAN
Good cleaning efficacy

Polysaccharide and blood test soil residues
Blocked irrigation system

Fibrin residues
Limited chemical cleaning efficiency for fibrin

Polysaccharide residues
Limited rinsing effect or bad water quality
PRODUCT GUIDES
Click to download: