TOSI®

Routine monitoring and validation of cleaning efficacy according to EN ISO 15883

  • A pre-prepared test giving efficient and repeatable results
  • User friendly and single use
  • Test soil correlated to human blood in actual practice
  • Transparent cover to enable visual check
  • The crevice simulates the joint of a surgical instrument
  • Clips for a quick and reproducible fixation

APPLICATION

TOSI test objects are simulated transparent surgical instruments. The contamination is correlating to the practical situation. If a cleaning process cannot clean the TOSI it is not efficient enough to clean blood contaminated instruments!

EVALUATION

The TOSI will not only give a yes/no answer for blood cleaning efficacy but will also give feedback regarding the cause of a problem:

CLEAN

Good cleaning efficacy

FIBRIN RESIDUE

Low chemical efficacy

HEMOGLOBIN

Low mechanical efficacy

Note: Contaminated instruments are a hygiene risk and need to be rejected. Instruments need to be free of any residue!: